Easy-Touch
(Obsolete)
The Easy-Touch has been replaced by the Easy-Touch Pro
128 points; expandable to 1,024 points in 128-point increments
Low Voltage Test
2 Wire
- Voltage: 4 V max
- Current: 3 μA to 6 mA
- Resistance: 0.1 Ω to 100 kΩ ± 1% ± 0.1 Ω, 100 kΩ to 5 MΩ ± 10%
4 Wire
- Voltage: 4 V max
- Current: 1 mA to 1 A
- Resistance: 0.001 Ω to 10 Ω ± 2% ± 0.001 Ω
Component Tests
- Diodes: Silicon, LEDs, zeners with breakdown > 4 V
- Resistors: 0.1 Ω to 100 kΩ ± 1% ± 0.1 Ω, 100 kΩ to 5 MΩ ± 10%
- Capacitors: 5 nF to 100 μF ±10% ± 0.02 nF
- Twisted Pairs: Verify proper pairing in twisted pair cables
High Voltage Test
Insulation Resistance Test
- Voltage: 50 to 1500 VDC ± 5% ± 5 V (2000 VDC optional)
- Resistance: 5 MΩ to 1000 MΩ ± 10%
Dielectric Withstand Test
- Voltage: 50 to 1500 VDC ± 5% ± 5 V (2000 VDC optional, 50 to 1000 VAC optional)
- Current Limit: 0.1 mA to 1.5 mA
- Max Capacitance Per Net: 150 nF @ 300 VDC, 90 nF @ 500 VDC, 45 nF @ 1000 VDC, 30 nF @ 1500 VDC, 9.5 nF @ 1000 VAC
- HV Energy Limit: 35 mJ
- HV Charge Limit: 45μC
Max Points Per Net
512
Digital Input/Output
4 Inputs / 6 Outputs, 24 V Open collector, +10 V and +5 V each current limited to 100 mA
Test Point Interface
Cirris Adapter System
User Interface
- Built-in Windows PC with Windows 7® Embedded
- Display: 10.4” Full-Color Graphical Touch Screen display
Power
115 / 230 V 50/60 Hz
Size
- Main unit: 17.5”w x 9”h x 5”d (44.5 cm x 23 cm x 13 cm)
- Add-on box: 6.5”w x 9”h x 5”d (16.5 cm x 23 cm x 13 cm)
Weight
- Main unit: 18 lbs (8.2 kg)
- Add-on box: 6.5 lbs (2.9kg)
Easy-Touch Voltage and Point Options
- 1500VDC 128 points (C151-ET) – Standard option
- Low Voltage 128 points (C15L-ET)
- 1500VDC 64 points (C1ET-64)
- 2000VDC 64 points (C2ET-64)
- 2000VDC 128 points (C200-ET)
S1KV-AC AC option for Easy-Touch (1000VAC)
- Adds an AC voltage source to the high voltage (hipot)
- Perform hipot tests that are in AC instead of DC
- Call 1-800-441-9910 for details on how to upgrade. We will need your serial number and unit model when you call.
SIO-EW Digital I/O option for Easy-Touch
- Remotely “Start Test” with an external switch, start button, input from a PLC, etc.
- Use Palm buttons, light curtains etc. as a hipot safety switch
SCPT-ET Scripting option for Easy-Touch
- Create complex tests that are easy for line workers to use
- Make highly customized reports or cable tags
- Control external devices such as lamps, hold-down clamps, markers, etc.
- Test devices such as switches, gas fuses, zener diodes, bi-color diodes, etc.
- Verify color and lighting of LEDs
- Display prompt messages
- Hipot different nets at different voltages
LAN-ET Server Software for Easy-Wire (up to 5 testers)
- Provides centralized management of Security Settings, Test Programs, and Test Reports data
- Manage your tester needs from any client connected to the Cirris Server Software
- Server and Clients must be running the same version
SLD-LK Slide-Lock Scanner Upgrade (New 1500VDC Easy-Touch testers ship with this upgrade)
- Allows operators to lock adapters in place without using a removable cover plate
- Hold adapters firmly in place to reduce connector wear
Performance Verification
- The Easy-Touch Performance Check Kit allows you to verify the calibration and proper operation of the Easy-Touch tester.
- For documentation on Performance Check Kits for the Easy-Touch click on the Documentation tab on this page.
Easy-Touch Version Lookup
On the main menu screen of Easy-Wire, the verson number is located on the mid-right side of the window.